Centrifuge SIGMA-step guide |
Author: Date:2009-05-17 |
Ⅰ. Operation panel
1. Start button: Start of centrifuges and started running. Conditions: The lid must be a good customs, the start button lights.
2. Stop button: Press the centrifuges will be the key to reducing the default speed and deceleration to a complete halt.
3. Key-off Cap: centrifuges used to open the lid. Conditions: complete cessation of centrifuges, the key lights off Cap.
4. Button: to choose or change the parameters.
II. Operation
1. Rotate button or click, the cursor appears. Its emergence is activated.
2. Rotary button so that the cursor moved to the need to revise the parameters, click the button to determine, SET will be light, that can be amended in the state.
3. Rotate button until you need to surface.
4. Click the button to confirm the figure.
5. If there are no parameters to modify or enter, the cursor automatically after 20 seconds out.
6. In the same way to change other parameters.
III. Notes
1. centrifuge should be smooth and well-ventilated table.
2. Apparatus around at least 30 cm of space.
3. Centrifugal time to trim, must be placed symmetrical.
4. Centrifugal pre-installation inspections must turn their back on the number of centrifuges and procedures set by the code are the same the other way round.
5. Every time after the end use will have to open the lid to allow centrifugal chamber to return to room temperature, and with a soft cloth to the centrifugal chamber, cooling water and stolen goods cleared.
6. Turn the screw shaft and to keep oil (SIGMA No NO70284), hanging baskets, and regular cleaning of oil adapter (SIGMA No NO70104).
7. Recommendations will be turned off every time spent alone on the refrigerator to preserve 4 degrees. |