There is a Temperature Resistor in the centrifuge chamber. It works as a temperature sensor. When the temperature in the chamber is higher than the OVERTEMP temperature, or when the temperature in is over 45 degrees, the system will cut off the power, at the same time, the OVERTEMP alert light will be light. Because the centrifuge only can work with a temperature which should be less than 38 degrees, if a centrifuge is running in a high speed, the temperature in the chamber will be much higher than 38 degrees, that means the cooling system is abnormal, and also the vacuum system.